In many industrial sectors , the airtightness of instruments and containers is of vital importance
For this reason, many scientific methods have been developed to detect any leaks.
The helium leak test is one of the most reliable and versatile tests for a number of reasons that we will explain in this article.
One of the most effective ways to test whether an ideally airtight container is actually leaking is to use tracer gases.
Tracer gases are a class of gases with specific characteristics that are introduced into the container and, by diffusing inside it, indicate the presence of openings.
Helium is particularly effective in this regard because it is light, odorless, colorless, and highly penetrating, capable of passing through even the smallest cracks, revealing potential leaks that would otherwise have gone unnoticed.
Discover the new series of Climatic Chambers for controlled climate testing.
Find out more.

Even if we don't think about it, we're surrounded by containers that need to be perfectly airtight to do their job.
For this reason, there are many sectors affected by Helium Leak Testing.
Here are the main ones.
Learn more: Discover our climatic chamber for testing lithium batteries (UN 38.3), designed and manufactured by FDM Environment Makers.
In order to properly perform the helium leak test, it is necessary to have an isolated, controlled and airtight environment.
For this, a thermo-vacuum chamber is often used , which we talked about here.
The sample to be tested is placed inside the chamber. Helium is then injected and the results are observed.
Using a specific instrument, usually a mass spectrometer (which we talked about here), any leaks are detected and located.
The specific procedures for performing helium leak testing are described in a number of international standards. The principal ones are ISO 20484:2017 and ASTM E498/E498M-11.
Build your environment directly online, based on your specific needs.
Find out more.

Like many other laboratory tests we have discussed on these pages, climatic chambers play a complementary but often fundamental role in helium leak testing
There are two functions that climatic chambers can perform in helium leak testing: conditioning and simulation.
Conditioning is when the sample to be analyzed must first be brought to a specific temperature to carry out the test in optimal conditions.
We speak of simulation when temperatures are altered during the test to simulate unfavorable environmental conditions that the sample will face during its life cycle (think of underwater pipes).
The FDM climate chamber is able to perform both tasks perfectly and complements helium leak testing and many other important product and material tests.
For further information, please do not hesitate to contact us.
Contact us for more information about this product.
Contact Us
Helium is the reference gas for leak testing for four reasons: (1) small atom (radius 31 pm) — passes through leaks that would stop other gases; (2) low atmospheric concentration (5.2 ppm) — low instrumental background noise; (3) inert — does not react with test materials; (4) non-toxic and non-flammable. Typical sensitivity of mass spectrometers : up to 10⁻¹² mbar·L/s (equivalent to one molecule/second).
MIL-STD-883 Method 1014 requires both: Fine Leak Test with helium to detect small leaks (10⁻⁸ ÷ 10⁻⁵ atm cc/s) — the component is pressurized in helium for 2-4 hours at 2-3 bar, then tested in an empty spectrometer. Gross Leak Test with Fluorinert FC-40 or PFE (C₈F₁₈) for large leaks (>10⁻⁴ atm cc/s) — the component is immersed in a volatile liquid at 85°C to generate visible bubbles. Without the gross test, a large leak would be missed because it would “saturate” the spectrometer too quickly.
MIL-STD-883 standard thresholds: Class S/B hermetic components < 5×10⁻⁸ atm cc/s He; commercial components < 10⁻⁷; pacemakers and implantable devices < 10⁻⁹ (ISO 5841). For fuel cells and batteries: < 10⁻⁶ to prevent moisture ingress. For cryogenic tanks: < 10⁻⁴ for safety. Limit selection is guided by expected service life and internal volume.
With calibrated reference leaks (sealed microporous metal capillaries, NIST/PTB certified): the detector reads the helium flow rate of the reference leak (typically 10⁻⁷ ÷ 10⁻⁴ atm cc/s) and applies a correction factor. Monthly primary calibration with a master reference; daily secondary calibration with an operational reference. The operator must be ASNT Level II qualified for leak testing according to ISO 9712. All tests are documented in reports with calibration traceability.
Do you need to pass a regulatory test or are you looking for a chamber designed to your test protocol? Tell us your needs: the FDM technical team will respond to you, without intermediaries.