Most climatic tests involve increasing the temperature and relative humidity to determine the reaction of the product or material being tested.
But what about the other side of the climate spectrum?
What are refrigeration systems and how many types of cooling chambers are there?
Today we are dealing with this topic.
In the industry, cooling chamber is an umbrella term that includes all those chambers that can bring and maintain the internal temperature at at least 0°.
For this reason, the function of a cooling chamber can be performed by at least three different types of laboratory instruments.
Discover the new series of Climatic Chambers for controlled climate testing.
Find out more.

Cooling chambers operate with two main cooling systems: disposable refrigerants and mechanical refrigeration systems. Let's look at them in detail.
Disposable refrigerants, as the name suggests, involve the use of refrigerants that are used only once in the cooling process and subsequently discarded. The most common laboratory refrigerants are ice and liquid nitrogen. They are primarily used for single tests where cooling is only required for a limited period or where ease of use and safety are paramount.
Mechanical refrigeration systems, on the other hand, are more similar to those used in common home refrigerators. They are the most common. The thermodynamic cycle uses a refrigerant to remove heat from an environment and cool it. The main components of a mechanical refrigeration system include the compressor, condenser, evaporator, and expander. The refrigerant is compressed, cooled, evaporated, and then expanded to create the cooling cycle.
The main use of cooling chambers is the storage of products and materials that must be kept at very low temperatures.
Three sectors are most involved: food, pharmaceuticals, and scientific research.
Walk-in refrigerators are mostly used in the food industry to store a large number of products, while laboratory refrigerators, more practical and compact, are used in the pharmaceutical industry and in scientific research institutes .
The main international standard that regulates the matter is ISO 23953 which provides guidelines on the classification, requirements and test conditions for countertop and non-countertop refrigerators used for professional purposes.
Build your environment directly online, based on your specific needs.
Find out more.

FDM – Environment Makers offers awide range of laboratory refrigerators and freezers capable of storing all the products and materials we have discussed.
For any further questions or concerns, please do not hesitate to contact us.
Contact us for more information about this product.
Contact Us
The F-Gas 2024/573 phase-down is eliminating gases with high GWP (Global Warming Potential). Current gases: R-449A, R-452A (HFC/HFO blend, GWP ~1400-1900, exiting); R-454C, R-455A (HFO blend low-GWP ~150-250, new standard); R-290 propane (GWP 3, natural, flammable A3 class 2 ATEX); R-744 CO₂ (GWP 1, high pressure 80+ bar, for ultra-freezers and commercial refrigeration); R-170 ethane and R-508B cascade for very low temperatures (-80°C and below).
With a two-stage cascade system : the first stage uses R-449A/R-404A to cool to -40°C, while the second stage uses R-508B (GWP 13,400 — outlet) or R-170 (natural ethane) to cool to -80°C. Modern systems use eco-friendly R-290/R-170 or R-744/R-170 cascades . Three cascade stages with R-14 (tetrafluoromethane) or Joule-Thomson expansion with helium gas are required for -150°C . Emerging alternatives include a refrigerant-free Stirling cryocooler for -186°C
Direct: evaporator inside the chamber, maximum heat exchange, lower cost, but risk of gas leaks in the chamber and local non-uniformity. Indirect (with glycol or brine): external exchanger + heat transfer fluid circuit in the chamber, greater uniformity (±0.3°C) and safety (no refrigerant inside), cost +20-30%. For ICH stability chambers and pharmaceutical applications, indirect is always preferred.
For systems containing ≥5 tonnes CO₂ equivalent (approximately 1.3 kg R-404A or 2.3 kg R-134a): (1) Annual leak check by an F-Gas certified technician; (2) Equipment register with traceability of loadings/rechargings; (3) Mandatory recovery at end of life; (4) Installer with F-Gas certification EU Reg 517/2014. Exceeding 500 t CO₂eq requires an automatic leak detection system.
Do you need to pass a regulatory test or are you looking for a chamber designed to your test protocol? Tell us your needs: the FDM technical team will respond to you, without intermediaries.